| Pressure Bar Graph: |
Range:-10 to 39.5cm H2O |
Resolution:0.5cmH2O |
| High Pressure Alarm: |
Range:5 to 99cm H2O,Off |
Resolution:1.0cmH2O |
| Low Pressure Alarm: |
Range:Off,1 to 20cm H2O |
Resolution:1.0cmH2O |
| Peak Pressure Display:(Ppk) |
Range:0 to 99.5cm H2O |
Resolution:0.5cmH2O |
| O2 concentration:(%) |
Range:6 to 100%. |
Resolution:0.5%O2 |
| High O2% Alarm: |
Range:19 to 100%,Off. |
Resolution:1.0%. |
| Low O2% Alarm: |
Range:18 to 99%. |
Resolution:1.0%. |
| O2 Sensor Response Time: |
90% in 7 seconds at room temperature |
|
| O2 Sensor Life: |
Over 1year under normal conditions. |
|
| Operating Temperature |
5ó to 45óC at 15 to 95% relative humidity |
|
| Storage Temperature: |
"-40ó to 60óC at 95%relativehumidity" |
|
| Atmospheric Pressure: |
600 to 1100 hPa |
|
| System Accuracy, Pressure: |
colSpan="2">± (1+3% of reading) rounded up to the nearest 0.5cm H2O |
|
| System Accuracy, Oxygen: |
±1% when calibrated to ambient pressure | |